GETINGE US SALES LLC: Medical Device Recall in 2019 - (Recall #: Z-0988-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Maquet XS Flat Screen Monitor Holder, PWD50+DF XS32 K3 SPE, Configuration Part Number ARD268400140A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Product Classification:
Class II
Date Initiated: January 8, 2019
Date Posted: March 13, 2019
Recall Number: Z-0988-2019
Event ID: 82074
Reason for Recall:
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Status: Terminated
Product Quantity: 6
Code Information:
Serial Numbers 1 2 3 4 5 6
Distribution Pattern:
The products were distributed to the following US states: AZ and TX.
Voluntary or Mandated:
Voluntary: Firm initiated