GETINGE US SALES LLC: Medical Device Recall in 2019 - (Recall #: Z-0989-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
Maquet XS Flat Screen Monitor Holder, PWD70+DF HD R XS32 K3 SPE, Configuration Part Number ARD268400150A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Class II
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Serial Numbers 500001 500002 500003 500004 500005 500006
The products were distributed to the following US states: AZ and TX.
Voluntary: Firm initiated