Getinge USA Inc: Medical Device Recall in 2012 - (Recall #: Z-2079-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128

Product Classification:

Class II

Date Initiated: May 25, 2012
Date Posted: August 1, 2012
Recall Number: Z-2079-2012
Event ID: 61974
Reason for Recall:

Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error.

Status: Terminated
Product Quantity: 8 washers
Code Information:

Serial numbers: W50024258; W50024259; W50026573; W50025790; W50017381; W50020296; W50028138; W50020952

Distribution Pattern:

Worldwide Distribution - USA including Maryland, Indiana, New Mexico, Massachuesetts, Texas, Kansas and the country of Canada

Voluntary or Mandated:

Voluntary: Firm initiated