Getinge USA Inc: Medical Device Recall in 2013 - (Recall #: Z-1994-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM --- Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623. Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.

Product Classification:

Class II

Date Initiated: August 2, 2013
Date Posted: August 28, 2013
Recall Number: Z-1994-2013
Event ID: 65866
Reason for Recall:

AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters.

Status: Terminated
Product Quantity: Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs)
Code Information:

LOT SR-430 (Expiration 02/06/2014)

Distribution Pattern:

Worldwide Distribution - USA and the country of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated