Getinge Usa Sales Inc: Medical Device Recall in 2024 - (Recall #: Z-2273-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322F5
Class II
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational failure.
Model Number/Part Number: 113322F5; UDI/DI: 04046768040052; Serial Numbers: 409 through 411, 413 through 416.
Domestic: KY, NC, NY, VA. International: Australia, Bulgaria, Croatia, Czech Republic, Denmark, France, Greece, Hong Kong, Ireland, Israel, Japan, Jordan, Kuwait, Mexico, New Zealand, Norway, Poland, Romania, Saudi Arabia, South Africa, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Voluntary: Firm initiated