Gk Photonics Inc: Medical Device Recall in 2019 - (Recall #: Z-2096-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Logolas Laser Projection Series
Product Classification:
Class II
Date Initiated: July 10, 2019
Date Posted: August 14, 2019
Recall Number: Z-2096-2019
Event ID: 83438
Reason for Recall:
The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.
Status: Ongoing
Product Quantity: N/A
Code Information:
Logolas Laser Projection Series
Distribution Pattern:
Unknown
Voluntary or Mandated:
FDA Mandated