Globus Medical, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2859-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
ALTERA Spacer ALTERA Spacer, 10x31, 9-13mm, 8¿
Product Classification:
Class II
Date Initiated: July 22, 2020
Date Posted: September 2, 2020
Recall Number: Z-2859-2020
Event ID: 86164
Reason for Recall:
Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Status: Terminated
Product Quantity: N/A
Code Information:
Part #: 1124.1112; Lot # / UDI #: BAX137PE / (01)00889095069686(10)BAX137PE; BAX167SE / (01)00889095069686(10)BAX167SE; BAX178FE / (01)00889095069686(10)BAX178FE; BAX208LE / (01)00889095069686(10)BAX208LE; BAX250CE / (01)00889095069686(10)BAX250CE; BAX348CE / (01)00889095069686(10)BAX348CE; BAX431NE / (01)00889095069686(10)BAX431NE
Distribution Pattern:
Worldwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated