Product Description:
ALTERA Spacer ALTERA Spacer, 10x31, 10-14mm, 8¿
Product Classification:
Class II
Date Initiated: July 22, 2020
Date Posted: September 2, 2020
Recall Number: Z-2860-2020
Event ID: 86164
Reason for Recall:
Specific lots of ALTERA¿ implants have internal components that may have been manufactured using stainless steel rather than the specified cobalt chromium molybdenum alloy.
Status: Terminated
Product Quantity: N/A
Code Information:
Part #: 1124.1113; Lot # / UDI #: BAX178LE / (01)00889095069693(10)BAX178LE; BAX178KE / (01)00889095069693(10)BAX178KE; BAX189PE / (01)00889095069693(10)BAX189PE; BAX348EE / (01)00889095069693(10)BAX348EE
Distribution Pattern:
Worldwide Distribution
Voluntary or Mandated: