GN Otometrics: Medical Device Recall in 2019 - (Recall #: Z-2553-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Gold Cup Leadwire and Snap Leadwire with the following component descriptions: 1. 24" Snap Leadwire, Blue; 2. 40" Snap Leadwire, Blue; 3. 24" Snap Leadwire, Red; 4. 40" Snap Leadwire, Red; 5. 24" Snap Leadwire, Green; 6. 40" Snap Leadwire, Green; 7. 24" Snap Leadwire, Brown; 8. 40" Snap Leadwire, Brown; 9. 24" Snap Leadwire, Orange; 10. 40" Snap Leadwire, Orange; 11. 24" Snap Leadwire, White; 12. 40" Snap Leadwire, White; 13. 24" Snap Leadwire, Black; 14. 40" Snap Leadwire, Black; 15. 24" Snap Leadwire, Yellow; 16. 40" Snap Leadwire, Yellow; 17. 24" Snap Leadwire, Set of 5; 18. 24" Snap Leadwire, Set of 7; 19. 40" Snap Leadwire, Set of 5; 20. 40" Snap Leadwire, Set of 7; 21. 24" Snap Leadwire, Set of 5; 22. 40" Snap Leadwire, Set of 5; 23. 48" Gold Cup Leadwires, Blue; 24. 48" Gold Cup Leadwires, Red; 25. 48" Gold Cup Leadwires, Green; 26. 48" Gold Cup Leadwires, Brown; 27. 48" Gold Cup Leadwires, Orange; 28. 48" Gold Cup Leadwires, White; 29. 48" Gold Cup Leadwires, Black; 30. 48" Gold Cup Leadwires, Yellow; 31. 48" Gold Cup Leadwires, Set of 5; 32. 48" Gold Cup Leadwires, Set of 8; Product Usage: The snap electrode leads are used in conjunction with disposable electrodes and the Gold Cup electrodes are reusable electrodes and leads. Both types of electrodes and leads are used for evoked potential testing for audiological, neurological and/or vestibular testing. The electrode leads are connected to the main device by the electrode lead connector and the electrode is attached to the patient contact end of electrode lead.

Product Classification:

Class II

Date Initiated: August 1, 2019
Date Posted: October 2, 2019
Recall Number: Z-2553-2019
Event ID: 83491
Reason for Recall:

The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm.

Status: Ongoing
Product Quantity: 4,882 units (41950 individual electrode leads supplied In 5,7 & 10 packs)
Code Information:

Part Numbers: 1. 7590310-24; 2. 7590310-40; 3. 7590311-24; 4. 7590311-40; 5. 7590312-24; 6. 7590312-40; 7. 7590313-24; 8. 7590313-40; 9. 7590314-24; 10. 7590314-40; 11. 7590315-24; 12. 7590315-40; 13. 7590316-24; 14. 7590316-40; 15. 7590317-24; 16. 7590317-40; 17. 7590318-24-5; 18. 7590318-24-7; 19. 7590318-40-5; 20. 7590318-40-7; 21. 7590319-24-5; 22. 7590319-40-5; 23. 7590440; 24. 7590441; 25. 7590442; 26. 7590443; 27. 7590444; 28. 7590445; 29. 7590446; 30. 7590447; 31. 7590450-5; 32. 7590450-8

Distribution Pattern:

Worldwide distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cote D'ivoire, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Korea, Republic of, Kuwait, Lebanon, Luxembourg, Macedonia, The Former Yugoslav, Malaysia, Maldives, Mayotte, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, Zimbabwe.

Voluntary or Mandated:

Voluntary: Firm initiated