Greatbatch Medical: Medical Device Recall in 2013 - (Recall #: Z-0751-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
45¿ Angle Driver / Z-H, Greatbatch Medical 10 000 Wahrie Drive, Clarance, NY 14031. Model #: P2661, P2662, T3635, T5586,T6031, T6164,T10161, T12628,T14879, T16422,T2487,T2713,T116057, T13635, MPU260045AO01, MPU260045ZH01. The Driver is intended to act as an interface between the cutting tool and a motorized driving system (i.e. surgical drill).
Class II
Greatbatch Medical has initiated a global voluntary recall and market withdrawal of its entire Small Angled Driver 45¿ product line. Greatbatch has received reports of noisy and / or rough operation when using these Small Angled Drivers.
All lots
Worldwide Distribution-USA including the states of CA, IN, KS, MA, NJ, PA, TN, and TX, and the countries of AUSTRALIA, BRAZIL, CANADA, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, PORTUGAL, and SWITZERLAND.
Voluntary: Firm initiated