Greatbatch Medical: Medical Device Recall in 2017 - (Recall #: Z-1190-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Offset Reamer Handle, T5766 Product Usage: The reamer handles are re-usable surgical instruments used in arthroplasty surgery of the hip joints, where either total joint replacement or resurfacing surgery is performed using a reamer cutting instrument. The reamer handles are intended for use with a surgical driver (active device). The reamer handle is connected at the distal end to a reamer. The T5766 reamer handle is an offset design with a drive shaft mechanism that spins on roller bearings and is encapsulated in an inner clam shell housing that is constrained within a Radel¿ tube and outer housing.

Product Classification:

Class II

Date Initiated: October 19, 2016
Date Posted: February 22, 2017
Recall Number: Z-1190-2017
Event ID: 76236
Reason for Recall:

On June 14, 2016, Greatbatch discovered a significant increase in the number of field complaints for the T5766 Offset Reamer Handle in which the U-joints of the drive chain component seizes during orthopedic procedures rendering the device non-functional. The 2016 failure rate for this failure exceeds the occurrence rate as defined in the DFMEA.

Status: Terminated
Product Quantity: 1,441
Code Information:

2481193, 2468380, 2599785, 2463309, 2468379, 2468398, 2468397, 2599516, 2671555, 2599517, 2670543, 2671558, 2734809, 2814828, 2814829, 2824496, 2858434, 2865404, 3030658, 2944397, 2977852, 2967214, 3000586, 3000590, 3030658, 3044689, 3071903, 3135658, 3088675, 3107504, 3188688, 3385844, 3188690, 3387046, 3397768.

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of IN and the countries of Canada, Austria, Germany, Japan.

Voluntary or Mandated:

Voluntary: Firm initiated