Product Description:

Haiou Needle Retractable Safety Syringe with Needle-1ML 23G x 1 (0.6x25mm). 200pcs/Box Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-23G25

Product Classification:

Class II

Date Initiated: June 10, 2022
Date Posted: August 17, 2022
Recall Number: Z-1507-2022
Event ID: 90224
Reason for Recall:

Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user

Status: Ongoing
Product Quantity: 7,218,000 units
Code Information:

Lot Number: 20JC2 Exp. Date: 2025/10/17; 21AA0 Exp Date: 2026/01/04

Distribution Pattern:

US Nationwide distribution in the state of CA.

Voluntary or Mandated:

Voluntary: Firm initiated