Haag-Streit USA Inc: Medical Device Recall in 2020 - (Recall #: Z-0697-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view

Product Classification:

Class II

Date Initiated: November 20, 2020
Date Posted: December 30, 2020
Recall Number: Z-0697-2021
Event ID: 86707
Reason for Recall:

Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient's eye

Status: Ongoing
Product Quantity: 3 units US
Code Information:

Floor stands FS 2-11, FS 2-15 with the software REF 615 588 in versions 2.0 to 3.3. FS 2-11 Ref. # 615H511 Serial Numbers:From 1001; FS 2-15 Ref. #615H515 Serial Numbers From 1001; HS Hi-R NEO 900 Ref. # 657 Serial Numbers 820 from 101; HS Hi-R NEO 900A Ref # 657 821 Serial Numbers from 101 Unique Device Identifier (UDI) HS Hi-R NEO 900 +EMWM657 820 0/$+serial no. HS Hi-R NEO 900A +EMWM657 821 0/$+serial no. FS 2 11 +EMWM 615H5110/$+serial no. FS 2-15 +EMWM 615H515 0/$+serial no.

Distribution Pattern:

US Nationwide distribution in the state of OK.

Voluntary or Mandated:

Voluntary: Firm initiated