Halyard Health, Inc: Medical Device Recall in 2015 - (Recall #: Z-2502-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
NovaPlus Trach Care* Closed Suction System for Adults, T-Piece, Directional Tip, 14F (Product Code 2V22059) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Product Classification:
Class II
Date Initiated: July 29, 2015
Date Posted: September 9, 2015
Recall Number: Z-2502-2015
Event ID: 71966
Reason for Recall:
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Status: Terminated
Product Quantity: N/A
Code Information:
Product Code 2V22059 - M5082T611
Distribution Pattern:
Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
Voluntary or Mandated:
Voluntary: Firm initiated