Halyard Health, Inc: Medical Device Recall in 2015 - (Recall #: Z-2507-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DSE, MDI (Product 8224); KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DSE (Product Code 22106); KimVent* Closed Suction System for Adults, 14 F, DSE, Wet Pak (Produce Code 221036) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Class II
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Product Code 8224 - AB5161U07; Product Code 22106 - AB5117U03, AB5124U08, AB5124U20, AB5131U03; Product Code 221036 - AB5138U15, AB5161U05
Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
Voluntary: Firm initiated