Halyard Health, Inc: Medical Device Recall in 2015 - (Recall #: Z-2510-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

NovaPlus Trach Care* Closed Suction System for Adults, Double Swivel Elbow, Directional Tip, 14F (Product Code 2V22109) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.

Product Classification:

Class II

Date Initiated: July 29, 2015
Date Posted: September 9, 2015
Recall Number: Z-2510-2015
Event ID: 71966
Reason for Recall:

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

Status: Terminated
Product Quantity: N/A
Code Information:

Product Code 2V22109 - M5076T629

Distribution Pattern:

Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada

Voluntary or Mandated:

Voluntary: Firm initiated