Hamilton Medical, Inc.: Medical Device Recall in 2015 - (Recall #: Z-2057-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.
Product Classification:
Class II
Date Initiated: March 24, 2015
Date Posted: August 5, 2015
Recall Number: Z-2057-2015
Event ID: 71234
Reason for Recall:
Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
Status: Terminated
Product Quantity: 248
Code Information:
Software version 2.20 and 2.41,
Distribution Pattern:
Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated