Hamilton Medical, Inc.: Medical Device Recall in 2016 - (Recall #: Z-0666-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator support to adults and pediatrics, and optionally infants and neonates. Intended areas of use include: MRI Dept., Intensive, intermediate, emergency and long-term acute care as well as transfer of patients within a hospital.
Product Classification:
Class II
Date Initiated: September 26, 2016
Date Posted: December 7, 2016
Recall Number: Z-0666-2017
Event ID: 75336
Reason for Recall:
Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.
Status: Terminated
Product Quantity: 44 units
Code Information:
Serial Numbers between 2001 and 2103
Distribution Pattern:
Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated