Hamilton Medical, Inc.: Medical Device Recall in 2023 - (Recall #: Z-2529-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Product Classification:
Class I
Date Initiated: July 26, 2023
Date Posted: September 27, 2023
Recall Number: Z-2529-2023
Event ID: 92873
Reason for Recall:
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Status: Ongoing
Product Quantity: 6 units
Code Information:
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Distribution Pattern:
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Voluntary or Mandated:
Voluntary: Firm initiated