Handicare Accessibility: Medical Device Recall in 2013 - (Recall #: Z-0171-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Handset component of the Handicare USA's Minivator Bath Bliss Reclining Bath Lifts (Bath Bliss), Part Number BP300-100/BP300-400RSUS (handset) - Model Numbers BBR0311USAB (blue) and BBR0311USSS (silver) Handset component devices operate the Minivator Bath Bliss Reclining Bath Lift, a non-AC-powered patient lift consisting of a motorized actuator, powered mechanical seating device, temporarily secured in a bath, and used to lift and lower a patient in the vertical or reclining position while bathing.

Product Classification:

Class II

Date Initiated: November 29, 2012
Date Posted: November 13, 2013
Recall Number: Z-0171-2014
Event ID: 63859
Reason for Recall:

Firm became aware that two handsets in the affected batches became inoperable while the user was in the tub.

Status: Terminated
Product Quantity: 3301 handsets
Code Information:

Handsets shipped with Bath Lifts belonging to the following serial numbers: BH05100001 - BH05129999 and BH06120001 - BH11129999; Handsets shipped as spare parts with the following serial numbers: BS05100001 - BS05129999 and BS06120001 - BS11129999.

Distribution Pattern:

Worldwide distribution: USA (Nationwide) and country of: Canada.

Voluntary or Mandated:

Voluntary: Firm initiated