Handicare Usa Inc: Medical Device Recall in 2017 - (Recall #: Z-0013-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
Product Classification:
Class II
Date Initiated: September 8, 2017
Date Posted: October 18, 2017
Recall Number: Z-0013-2018
Event ID: 78096
Reason for Recall:
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
Status: Terminated
Product Quantity: 8 units
Code Information:
A6M0001526, A6M0001527, A6M0001528, A6M0001529, A6M0001530, A6M0001485, A6M0001486, A6M0001498
Distribution Pattern:
Us and Canada
Voluntary or Mandated:
Voluntary: Firm initiated