Hansen Medical Inc: Medical Device Recall in 2015 - (Recall #: Z-1449-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.
Product Classification:
Class II
Date Initiated: March 18, 2015
Date Posted: April 22, 2015
Recall Number: Z-1449-2015
Event ID: 70847
Reason for Recall:
Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.
Status: Terminated
Product Quantity: 19 total installed systems
Code Information:
All serial numbers.
Distribution Pattern:
Worldwide Distribution - US Distribution and the countries of : Saudi Arabia, Australia, France, Italy, Germany, and United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated