HeartSine Technologies Ltd: Medical Device Recall in 2024 - (Recall #: Z-1820-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P
Class II
Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.
UDI-DI: 5060167125379, 5060167120688, 5060167125720, 5060167128066, 5060167122330, 5060167122491, M727SAM450P, 5060167122514, 5060167121371, 5060167120671, 5060167121630, 5060167128042, 5060167120701, 5060167128080, 5060167120695, 5060167120725, 5060167120718, 5060167120909, 5060167120916, 5060167122071, 5060167120893. Devices affected by this notification begin with the following prefixes and device codes: 16D, 16E, 16G, 17D, 17E, 17G, 18D, 18E, 18G, 19D, 19E, 19G, 20D, 20E, 20G, 21D, 21E, 21G, 22D, 22E, 22G, 23D, 23E, 23G, 24D, 24E, 24G Serial Numbers: The prefix (device identifier) consists of the manufacturing date (YY) and the device model (B, D, E, G, or H) and an 8-digit serial number string. Example: 16D00001234
US Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GC, HI, IA, ID, IL, IN, KS, KY, LA, MA, MB, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY.
Voluntary: Firm initiated