HeartSine Technologies Ltd: Medical Device Recall in 2024 - (Recall #: Z-1820-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

HeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450P

Product Classification:

Class II

Date Initiated: April 8, 2024
Date Posted: May 22, 2024
Recall Number: Z-1820-2024
Event ID: 94076
Reason for Recall:

Automated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user during use of the device. Visual instructional icons will still be present, but this issue could potentially lead to no therapy or a delay in therapy.

Status: Ongoing
Product Quantity: 183,248
Code Information:

UDI-DI: 5060167125379, 5060167120688, 5060167125720, 5060167128066, 5060167122330, 5060167122491, M727SAM450P, 5060167122514, 5060167121371, 5060167120671, 5060167121630, 5060167128042, 5060167120701, 5060167128080, 5060167120695, 5060167120725, 5060167120718, 5060167120909, 5060167120916, 5060167122071, 5060167120893. Devices affected by this notification begin with the following prefixes and device codes: 16D, 16E, 16G, 17D, 17E, 17G, 18D, 18E, 18G, 19D, 19E, 19G, 20D, 20E, 20G, 21D, 21E, 21G, 22D, 22E, 22G, 23D, 23E, 23G, 24D, 24E, 24G Serial Numbers: The prefix (device identifier) consists of the manufacturing date (YY) and the device model (B, D, E, G, or H) and an 8-digit serial number string. Example: 16D00001234

Distribution Pattern:

US Nationwide distribution including in the states of AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GC, HI, IA, ID, IL, IN, KS, KY, LA, MA, MB, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY.

Voluntary or Mandated:

Voluntary: Firm initiated