Hill-Rom, Inc.: Medical Device Recall in 2017 - (Recall #: Z-1816-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale Upgrade Kits, Model Numbers P1947A and P1948A).

Product Classification:

Class II

Date Initiated: February 23, 2017
Date Posted: April 19, 2017
Recall Number: Z-1816-2017
Event ID: 76506
Reason for Recall:

The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.

Status: Terminated
Product Quantity: 3037
Code Information:

P1900 SN: K328AM6065 to Q245AM0691 PR1900 SN: L082AL7316 to R343AL4516 U1900 SN: L257AM0541 to O311AM7169 Scale Upgrade Kit P1957A produced after March 13, 2013 (not serialized).

Distribution Pattern:

US Nationwide. AUSTRALIA BANGLADESH BRAZIL CANADA CHINA GERMANY INDIA ITALY KUWAIT PAKISTAN SAUDI ARABIA THAILAND UNITED ARAB EMIRATES CYPRUS GREECE NETHERLANDS

Voluntary or Mandated:

Voluntary: Firm initiated