Hobbs Medical, Inc.: Medical Device Recall in 2024 - (Recall #: Z-0470-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4706
Product Classification:
Class II
Date Initiated: October 4, 2024
Date Posted: November 27, 2024
Recall Number: Z-0470-2025
Event ID: 95577
Reason for Recall:
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
Status: Ongoing
Product Quantity: 6 units
Code Information:
Catalog Number: 4706; UDI-DI: M84947060; Lot Number: SD09-23-089.
Distribution Pattern:
Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Voluntary or Mandated:
Voluntary: Firm initiated