Horiba Instruments Inc: Medical Device Recall in 2016 - (Recall #: Z-0740-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.
Class II
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
All Lots
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
Voluntary: Firm initiated