Horiba Instruments Inc: Medical Device Recall in 2016 - (Recall #: Z-0742-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
ABX PENTRA Triglycerides CP ref. A11A01640 is an in vitro diagnostic assay for the quantitative determination of triglycerides in human serum and plasma based on an enzymatic colorimetric test. It is composed of a 99 ml mono-reagent cassette. Reagent is a chemical solution with additives.
Class II
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
All Lots
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
Voluntary: Firm initiated