Hospira Inc.: Medical Device Recall in 2013 - (Recall #: Z-0205-2014)
See the recall detail below. You can also see other recalls from the same firm in 2013.
Plum A+ Infusion Pump; List Number: 11973; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
Class II
Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
List No. 11973; Serial Numbers: 12570148 , 12570150 , 12570154 , 12570163 , 12570170 , 12570177 , 12570204 , 12570211 , 12570212 , 12570255 , 12570270 , 12570319 , 12570359 , 12570385 , 12570413 , 12570425 , 12570455 , 12570487 , 12570662 , 12570695 , 12570854 , 12570915 , 12571080 , 12571243 , 12571437 , 12571631 , 12571692 , 12571776 , 12571826 , 12571835 , 12571837 , 12571842 , 12571877 , 12571880 , 12571886 , 12572317 , 12572369 , 12572586 , 12572909 , 12573547 , 12573567 , 12573837 , 12573990 , 12574424 , 12574501 , 12574524 , 12574556 , 12574599 , 12574619 , 12574731 , 12574749 , 12574853 , 12574858 , 12574860 , 12574865 , 12574961 , 12575626 , 12575698 , 12575886 , 12576125 , 12576443 , 12576485 , 12576518 , 12576593 , 12576609 , 12576710 , 12576722 , 12577271 , 12577685 , 12578856 , 12579183
Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Croatia, Dominican Republic, Turks and Caicos.
Voluntary: Firm initiated