Hospira Inc.: Medical Device Recall in 2013 - (Recall #: Z-0625-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software; Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system that provides precise delivery of multiple i.v. therapies across the general spectrum of clinical care, including ICU/CCU, OR, PACU, ER, Med/Surg, Pediatrics, Outpatient clinics and Home care

Product Classification:

Class II

Date Initiated: December 7, 2012
Date Posted: January 9, 2013
Recall Number: Z-0625-2013
Event ID: 63772
Reason for Recall:

The door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.

Status: Terminated
Product Quantity: 28,599 units
Code Information:

list numbers 12348, 12618, and 20678; all serial numbers

Distribution Pattern:

Worldwide Distribution -- USA, including Puerto Rico, and the countries of Austria, Australia, Argentina, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, Finland, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates, and Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated