Human Care USA, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0851-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Hanger-Bar 2P 45cm

Product Classification:

Class II

Date Initiated: November 1, 2021
Date Posted: April 6, 2022
Recall Number: Z-0851-2022
Event ID: 89509
Reason for Recall:

The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.

Status: Ongoing
Product Quantity: 10 devices
Code Information:

Part Number 72754 UPC Code: 881608800397 EAN Code: 7331599999707 Lot Number: 33218501

Distribution Pattern:

U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.

Voluntary or Mandated:

Voluntary: Firm initiated