Hummingbird Med: Medical Device Recall in 2016 - (Recall #: Z-1636-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used for blood transfer and collection from Peripheral Arterial Line Catheters.

Product Classification:

Class I

Date Initiated: October 1, 2015
Date Posted: June 1, 2016
Recall Number: Z-1636-2016
Event ID: 72350
Reason for Recall:

Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.

Status: Terminated
Product Quantity: 11500 units
Code Information:

Lot 15180, 15286, 15287, 15300, 15305

Distribution Pattern:

Distributed in the states of CA, KY, MD, GA, and IL.

Voluntary or Mandated:

Voluntary: Firm initiated