I-Flow Corporation: Medical Device Recall in 2012 - (Recall #: Z-2253-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

ON-Q pump with ONDEMAND bolus button, ON-Q C-bloc with ONDEMAND; ON-Q C-bloc with Select a Flow and ONDEMAND Model/Reference #: 7000180, 7000181, 7000182, 7000367, 7000397, CB003, CB005, CB006, P400X1-7+5/60, P400X2-14+5/30, P400X5+5/60, and PMB01. Part Numbers: 101347100, 101347300, 101347400, 101360200, 101368400, 101368600, 101368700, 102987000, 102987200, 102987300, 103161800, 103534300, 5001438, 5001470, 5001651, 5001780, 5001808, 5001809, 5001812, 7000180, 7000181, 7000182, & 7000367. Indicated for continuous and or intermittent delivery of medication to or around surgical wound sites and or close proximity to nerves.

Product Classification:

Class I

Date Initiated: May 8, 2012
Date Posted: September 5, 2012
Recall Number: Z-2253-2012
Event ID: 62769
Reason for Recall:

It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected.

Status: Terminated
Product Quantity: 80,646 units total (65,926 units in US)
Code Information:

Products having a 10-digit Lot # that is greater than 0200521454 (i.e. 0200521455 and greater) are not impacted by this recall.

Distribution Pattern:

Worldwide Distribution -- USA, Canada, Australia, the EU (Germany), United Arab Emirates, and Cyprus.

Voluntary or Mandated:

Voluntary: Firm initiated