Iba Dosimetry: Medical Device Recall in 2017 - (Recall #: Z-2220-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T

Product Classification:

Class II

Date Initiated: April 26, 2017
Date Posted: June 14, 2017
Recall Number: Z-2220-2017
Event ID: 77301
Reason for Recall:

When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.

Status: Terminated
Product Quantity: 110 units
Code Information:

DS50-001-XL-T, Serial Numbers: 24887, 25084, 25085 DS50-002-XL-T, Serial Numbers: 24185, 25219, 25470, 24073, 24187, 24480, 24888, 25089, 24889, 24184, 24813, 24814, 24866, 24867, 24890, 25088, 25220, 24482, 24186, 25090, 24626, 24074, 24189, 25218, 25468, 25469 DS50-003-XL-T, Serial Numbers: 24437, 24260, 24639, 25472, 24257, 24291, 24630, 24640, 24868, 25093, 25478, 24180, 24182, 24259, 24323, 24439, 25222, 24179, 24290, 24293, 24633, 24637, 25091, 25223, 24627, 24077, 24634, 24438, 25476, 25221, 24292, 24263, 24322, 24435, 24628, 24642, 25094, 25479, 24177, 24178, 24641, 25481, 24261, 24076, 24871, 25224, 24078, 24079, 24080, 24175, 24176, 24183, 24256, 24258, 24262, 24264, 24265, 24289, 24321, 24436, 24629, 24631, 24635, 24636, 24638, 24869, 24891, 24892, 24893, 24894, 24896, 24897, 25092, 25095, 25225, 25477 DS50-102-XL-T, Serial Numbers: 24190, 24481, 25467 DS50-103-XL-T, Serial Numbers: 24870, 24898

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated