ICU Medical de Mexico, S.A. de C.V.: Medical Device Recall in 2018 - (Recall #: Z-2803-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

18 cm (7") PUR Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, NanoClave, Rotating Luer.

Product Classification:

Class II

Date Initiated: March 1, 2018
Date Posted: August 22, 2018
Recall Number: Z-2803-2018
Event ID: 80460
Reason for Recall:

ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.

Status: Terminated
Product Quantity: 100 pouches
Code Information:

Item Number 011-AM6111. a. Lot Number 3442842, UDI Number (01) 1 0840619 08918 1 (17) 220501 (30) 50 (10) 3442842, Expiration Date 05/01/2021; b. Lot Number 3508246, UDI Number (01) 1 0840619 08918 1 (17) 220801 (30) 50 (10)3508246, Expiration Date 08/01/2022

Distribution Pattern:

US distribution to AZ, MA, NC, OH, TX, VA, WA. Worldwide distribution to Belgium, France, Germany.

Voluntary or Mandated:

Voluntary: Firm initiated