ICU Medical, Inc.: Medical Device Recall in 2016 - (Recall #: Z-1365-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
Diana Automated Compounding System, Diana Onco Plus 1.2, Item No. ASN222, ASN222-GST, ASN222-R The Diana is an automated compounding device for the accurate, safe and efficient preparation of solutions by a pharmacist prior to delivery for infusion into patient.
Class II
ICU Medical Inc. has received a report of compounded volumes not meeting the accuracy specification by the Diana Compounding System when mixing volumes under 5ml.
All Serial Numbers
Nationwide Distribution to Nationwide in AL, AZ, CA, CO, CT, DC, GA, IL, MD, MI, MN, MO, NC, NH, NJ, NY, OH, OK, PA, TN, TX, VA.. Worldwide: AU, BE, BR, CA, CH, DE, DZ, EE, EG, ES, FI, FR, HK, HR, IL, IT, JO, KR, MX, NL, PL, SA, SE, SI, TR.
Voluntary: Firm initiated