ICU Medical, Inc.: Medical Device Recall in 2019 - (Recall #: Z-1095-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
icumedical Oncology Kit w/12" (30cm) Ext Set w/Spiros w/Red Cap, Clamp, Graduated Adapter; ChemoClave Vented Vial Spike, 20mm; ChemoClave Vented Vial Spike, 13mm; Spiros wRed Cap, REF 3944. Used in IV administration sets. The ChemoCLAVE Cytotoxic Medication Preparation and Delivery System consists of 6 previously cleared components (CLAVE, Spikes, SPIROS, GENIE, Vial Access, and Admin Sets) that can be combined into various configurations intended for use in the preparation and patient administration of cytotoxic medications.
Class I
There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.
Lot Numbers: 3846116, 3855565
Worldwide distribution: US nationwide and countries of: Canada, France, and Singapore.
Voluntary: Firm initiated