ICU Medical, Inc.: Medical Device Recall in 2020 - (Recall #: Z-0702-2021)
See the recall detail below. You can also see other recalls from the same firm in 2020.
30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10; Spinning Spiros Closed Male Luer, Red Cap, REF: CH2000S-C; Spiros, Closed Male Luer w/Red Cap, 10 Units, RES: CH2000SC-10; Spiros, Closed Male Luer w/Red Cap, 5 Units, REF: CH2000SC-5; Spinning Spiros Closed Male Luer, Purple Cap, REF: CH2000S-PC; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC; Spiros, Purple Cap, Drop-In Priming Cap, REF: CH2000S-PCP; Spinning Spiros Closed Male Luer w/Red Cap, REF: CH3515; Spinning Spiros Closed Male Luer, REF: Z7072
Class II
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Lots: 4749745, 4764622, 4790769, 4806298, 4865663, 4712177, 4749754, 4757981, 4764633, 4793559, 4903194, 4726692, 4756786, 4800185, 4895432, 4726695, 4749730, 4757614, 4761878, 4784740, 4794303, 4895435, 4713514, 4713516, 4749750, 4749755, 4750947, 4750950, 4750952, 4757618, 4763178, 4806832, 4806838, 4849195, 4858381, 4895446, 4711932, 4726709, 4756789, 4761899, 4783763, 4858387, 4962299, 4726711, 4749767, 4774705, 4784757, 4795443, 4857778, 4727446, 4749771, 4711937, 4727450, 4882569, 4741999, 4757713, 4793544, 4908662, 4749745, 4764622, 4790769, 4865663, 4901210, 4749754, 4764633, 4603678, 4726692, 4756786, 4800185, 4895432, 4713516, 4749733, 4749750, 4749755, 4750947, 4750952, 4763174, 4763178, 4774701, 4774702, 4806838, 4849192, 4849195, 4857774, 4858382, 4895442, 4895446, 4896004, 4902104, 4726709, 4761899, 4727446, 4749771, 4774000, 4849201, 4877398, 4877399
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Voluntary: Firm initiated