ICU Medical, Inc.: Medical Device Recall in 2020 - (Recall #: Z-0702-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger, REF: 20124-01; Spinning Spiros Closed Male Luer, Red Cap, REF: 20130-01; Spinning Spiros, Closed Male Luer, REF: CH2000S; Spinning Spiros Closed Male Luer, 10 units, REF: CH2000S-10; Spinning Spiros Closed Male Luer, Red Cap, REF: CH2000S-C; Spiros, Closed Male Luer w/Red Cap, 10 Units, RES: CH2000SC-10; Spiros, Closed Male Luer w/Red Cap, 5 Units, REF: CH2000SC-5; Spinning Spiros Closed Male Luer, Purple Cap, REF: CH2000S-PC; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC; Spiros, Purple Cap, Drop-In Priming Cap, REF: CH2000S-PCP; Spinning Spiros Closed Male Luer w/Red Cap, REF: CH3515; Spinning Spiros Closed Male Luer, REF: Z7072

Product Classification:

Class II

Date Initiated: August 31, 2020
Date Posted: December 30, 2020
Recall Number: Z-0702-2021
Event ID: 86426
Reason for Recall:

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Status: Ongoing
Product Quantity: 1856578
Code Information:

Lots: 4749745, 4764622, 4790769, 4806298, 4865663, 4712177, 4749754, 4757981, 4764633, 4793559, 4903194, 4726692, 4756786, 4800185, 4895432, 4726695, 4749730, 4757614, 4761878, 4784740, 4794303, 4895435, 4713514, 4713516, 4749750, 4749755, 4750947, 4750950, 4750952, 4757618, 4763178, 4806832, 4806838, 4849195, 4858381, 4895446, 4711932, 4726709, 4756789, 4761899, 4783763, 4858387, 4962299, 4726711, 4749767, 4774705, 4784757, 4795443, 4857778, 4727446, 4749771, 4711937, 4727450, 4882569, 4741999, 4757713, 4793544, 4908662, 4749745, 4764622, 4790769, 4865663, 4901210, 4749754, 4764633, 4603678, 4726692, 4756786, 4800185, 4895432, 4713516, 4749733, 4749750, 4749755, 4750947, 4750952, 4763174, 4763178, 4774701, 4774702, 4806838, 4849192, 4849195, 4857774, 4858382, 4895442, 4895446, 4896004, 4902104, 4726709, 4761899, 4727446, 4749771, 4774000, 4849201, 4877398, 4877399

Distribution Pattern:

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Voluntary or Mandated:

Voluntary: Firm initiated