ICU Medical, Inc.: Medical Device Recall in 2020 - (Recall #: Z-1911-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

REF: AH7117 111" 10 Drop Admin Set w/2 Inj Sites, Rotating Luer, Sterile R, For Veterinary Use Only, (01) 1 0887709 051475

Product Classification:

Class II

Date Initiated: October 4, 2019
Date Posted: May 13, 2020
Recall Number: Z-1911-2020
Event ID: 85053
Reason for Recall:

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

Status: Terminated
Product Quantity: 1175 units
Code Information:

Lot number 4173657 *Product not distributed within the US

Distribution Pattern:

US:AK,AL,AR,AZ,CA,CO,CT,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,SD,TN,TX,UT,VA,WA,WI,WY OUS: Canada *** "REF" numbers AH7117 and B30211 were not marketed or distributed within the US***

Voluntary or Mandated:

Voluntary: Firm initiated