ICU Medical, Inc.: Medical Device Recall in 2021 - (Recall #: Z-1565-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
REF CH3397 Oncology Kit w/Spiros¿, Red Cap, Bag Spike w/Clave Additive Port LOT 5118610 - Product Usage: for use in the preparation and patient administration of cytotoxic medications.
Class II
Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
REF: CH3397 UDI: (01) 1 0840619 07367 8 (17) 251201 (30) 50 (10) 5118610 Lot Number: 5118610
Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA and WA. The countries of Belgium, Canada, France, Italy, South Africa, Spain, United Kingdom.
Voluntary: Firm initiated