ICU Medical, Inc.: Medical Device Recall in 2021 - (Recall #: Z-1567-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

REF 20124-01 30" (76 cm) Appx 3.2 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Bag Hanger LOT 5142971 - Product Usage: provides access for the administration of fluids from a container to a patients vascular system through the administration sets needle or catheter (which is inserted into a vein).

Product Classification:

Class II

Date Initiated: March 16, 2021
Date Posted: May 19, 2021
Recall Number: Z-1567-2021
Event ID: 87703
Reason for Recall:

Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.

Status: Ongoing
Product Quantity:
Code Information:

REF: 20124-01 UDI: (01) 1 0887709 07925 7 (17) 260101 (30) 50 (10) 5125197 Lot Number: 5142971

Distribution Pattern:

Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA and WA. The countries of Belgium, Canada, France, Italy, South Africa, Spain, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated