ICU Medical Inc: Medical Device Recall in 2023 - (Recall #: Z-1566-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Replacement Battery List Number SUB0000594 found in Plum A+ & Plum A+3 Infusion Systems, List Numbers: a) 11005, b) 11971, c) 12391, d) 12618, e) 20678, f) 20679, g) 20792, h) 60629, i) 12348, j) 11973
Product Classification:
Class I
Date Initiated: March 22, 2023
Date Posted: May 24, 2023
Recall Number: Z-1566-2023
Event ID: 91874
Reason for Recall:
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Status: Ongoing
Product Quantity: 23815 units
Code Information:
All devices containing CSB BATTERY displaying the first two (2) characters as 22 or lower
Distribution Pattern:
worldwide
Voluntary or Mandated:
Voluntary: Firm initiated