ICU Medical Inc: Medical Device Recall in 2023 - (Recall #: Z-1567-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
ASM Replacement Battery, Component Number SUB0000864
Product Classification:
Class I
Date Initiated: March 22, 2023
Date Posted: May 24, 2023
Recall Number: Z-1567-2023
Event ID: 91874
Reason for Recall:
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
Status: Ongoing
Product Quantity: 9984 units
Code Information:
Lot Numbers: 6036488, 6827738, 6827739, 6827740, 8408750, 8408755, 8408757
Distribution Pattern:
worldwide
Voluntary or Mandated:
Voluntary: Firm initiated