ICU Medical, Inc.: Medical Device Recall in 2025 - (Recall #: Z-0154-2026)
See the recall detail below. You can also see other recalls from the same firm in 2025.
IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014
Class I
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
REF/UDI-DI/Lot: B33359/10887709007823/14070544, 14126963; B9213/10840619045859/14130197; B9732/10887709015033/14149636; B9733/10887709015040/14176779; 011-C7014/10840619005068/14029638, 14065820, 14110075, 14211300, 14129823
Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
Voluntary: Firm initiated