ICU Medical, Inc.: Medical Device Recall in 2025 - (Recall #: Z-1991-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

ICU Medical Plum Solos, Item number 400011001; infusion pump

Product Classification:

Class II

Date Initiated: April 22, 2025
Date Posted: June 25, 2025
Recall Number: Z-1991-2025
Event ID: 96777
Reason for Recall:

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

Status: Ongoing
Product Quantity: 214 units
Code Information:

UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173, 60000174, 60000176, 60000177, 60000178, 60000179, 60000180, 60000183, 60000187, 60000189, 60000190, 60000191, 60000192, 60000198, 60000199, 60000201, 60000203, 60000204, 60000205, 60000206, 60000207, 60000208, 60000209, 60000210, 60000212, 60000213, 60000214, 60000215, 60000216, 60000217, 60000223, 60000224, 60000225, 60000226, 60000227, 60000228, 60000229, 60000230, 60000234, 60000236, 60000237, 60000238, 60000239, 60000240, 60000241, 60000242, 60000244, 60000245, 60000246, 60000247, 60000249, 60000251, 60000252, 60000255, 60000257, 60000259, 60000263, 60000264, 60000265, 60000266, 60000268, 60000269, 60000271, 60000273, 60000274, 60000275, 60000276, 60000277, 60000279, 60000280, 60000281, 60000282, 60000283, 60000284, 60000285, 60000286, 60000288, 60000291, 60000292, 60000295, 60000296, 60000297, 60000298, 60000300, 60000155, 60000193, 60000197, 60000211, 60000243, 60000270, 60000272, 60000123, 60000168, 60000175, 60000195, 60000221, 60000256, 60000294.

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

Voluntary or Mandated:

Voluntary: Firm initiated