Imaging Sciences International, LLC: Medical Device Recall in 2012 - (Recall #: Z-2463-2012)
See the recall detail below. You can also see other recalls from the same firm in 2012.
DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System
Class II
A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)
Catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3
Worldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.
Voluntary: Firm initiated