Immuno-Mycologics, Inc: Medical Device Recall in 2017 - (Recall #: Z-1109-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Product Classification:
Class II
Date Initiated: December 27, 2016
Date Posted: February 8, 2017
Recall Number: Z-1109-2017
Event ID: 76159
Reason for Recall:
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
Status: Terminated
Product Quantity: 630 units
Code Information:
161028133653JH, 161003111544KH, and 161007105036JG
Distribution Pattern:
Nationwide throughout the US
Voluntary or Mandated:
Voluntary: Firm initiated