In2bones USA, LLC: Medical Device Recall in 2019 - (Recall #: Z-0670-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
NeoSpan SuperElastic Compression Staple w/Instruments, Size 10x10x10, Part Number T50 SN010
Product Classification:
Class II
Date Initiated: November 14, 2018
Date Posted: January 2, 2019
Recall Number: Z-0670-2019
Event ID: 81655
Reason for Recall:
The NeoSpan Compression Staple (10x10x10) kits from two lots contain the wrong size drill bit. They contain a 3mm drill bit instead of a 2mm drill bit.
Status: Terminated
Product Quantity: 30 units
Code Information:
Lot Number: 9G88C0
Distribution Pattern:
TX, VA, PA, MO, NJ, FL, TN, NC, OH, WI, LA
Voluntary or Mandated:
Voluntary: Firm initiated