INCIPIO DEVICES: Medical Device Recall in 2024 - (Recall #: Z-0549-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Offset Cup Impactor - 3/8-16 interface. Reusable surgical instrument used to position and place acetabular sup implants.
Product Classification:
Class II
Date Initiated: October 30, 2024
Date Posted: December 4, 2024
Recall Number: Z-0549-2025
Event ID: 95653
Reason for Recall:
Recalled lot does not meet specification and has reported incidents of fracture of the TIG weld.
Status: Ongoing
Product Quantity: 100 units
Code Information:
Product Reference No.: 50341105; UDI-DI: 07630055304735; Lot No.: 24A0087.
Distribution Pattern:
US Nationwide distribution in the state of TN.
Voluntary or Mandated:
Voluntary: Firm initiated