Incisive Surgical Inc: Medical Device Recall in 2015 - (Recall #: Z-1545-2015)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Insorb Subcuticular Skin Stapler. INSORB staples are made from an absorbable copolymer which is a synthetic polyester derived from lactic and glycolic acids. It is chemically similar to other surgical glycolide/lactide-based copolymers. Polyglycolic/polylactic acid copolymers degrade in vivo by hydrolysis to glycolic acid and lactic acid which are then absorbed and metabolized by the body.
Product Classification:
Class II
Date Initiated: March 19, 2015
Date Posted: May 6, 2015
Recall Number: Z-1545-2015
Event ID: 70906
Reason for Recall:
A previous Turkish distributor over-labeled a portion of the lot with an expiration date of 2016-05. The expiration date at the time of manufacturing was 2014-05.
Status: Terminated
Product Quantity: 210 cartons (containing 6 INSORB staplers per carton)
Code Information:
Lot 121901
Distribution Pattern:
International Distribution Only - Turkey.
Voluntary or Mandated:
Voluntary: Firm initiated