Infusion Pump Repair: Medical Device Recall in 2021 - (Recall #: Z-1544-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Class I
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Devices distributed between 01/15/2021 and 03/14/2021. Sale Order #/Serial #: 1599930872/ 12335834, 13617863, 13447266, 13089105, 12333728, 13616451, 13618575, 12729374; 82943/ 4122252, 4122320; 01112101A/ 12638236, 13813590, 12639090, 13744738, 9951014
California
Voluntary: Firm initiated